/ 제품 / Laminar Flow and Containment Equipment 에 대한 리뷰 보기 / Class II Type A2 Biological Safety Cabinet
Laminar Flow and Containment Equipment 에 대한 리뷰 보기

Class II Type A 2 생물학적 안전 캐비닛

A recirculating Class II cabinet that combines inward face airflow, HEPA-filtered vertical downflow and filtered exhaust for three-way protection.
  • Personnel Product and Environmental Protection
  • HEPA-Filtered Downflow
  • Negative-Pressure Contaminated Plenums
  • Recirculating A2 Airflow
견적 요청 제품 찾아보기

상품 소개

The Type A2 cabinet draws room air inward at the front grille while HEPA-filtered downflow protects the work zone. A portion of the filtered air is recirculated and the balance is exhausted through a HEPA stage, either into the room or through an appropriate canopy connection. Contaminated plenums are maintained under negative pressure or surrounded by negative-pressure sections. The cabinet is selected and installed for biological particulate work; volatile chemical use requires a separately evaluated exhaust strategy.

주요 기능

The front inflow grille captures room air before it crosses the work surface.

Vertical filtered downflow limits cross-contamination across the product zone.

Negative-pressure plenum design provides a fail-safe boundary around contaminated air paths.

A motorized or counterbalanced sash maintains the selected operating height.

Independent inflow and downflow sensing supports real-time airflow monitoring.

고객 제품 관심사

A Type A2 cabinet must balance operator capture and clean product downflow while keeping recirculated air paths safely enclosed.

Hands crossing the front grille disrupt containment.A properly sized grille and stable inflow maintain capture around normal arm movement.
Items block downflow and create cross-contamination zones.A wide supply diffuser and organized work layout preserve first-air paths.
Filter loading shifts the inflow/downflow balance.Separate airflow sensors and automatic blower control maintain the selected operating range.
The cabinet is used with unsuitable volatile chemicals.Application guidance distinguishes A2 biological particulate work from total-exhaust chemical handling.

일반적인 응용 프로그램

Cell cultureClinical microbiologyNon-volatile biological sample preparationSterile biological product handling

기술 매개변수

Typical configurable construction ranges for a Class II Type A2 biological safety cabinet are listed below.

핵심 재료SUS304 single-piece work-zone liner with powder-coated steel outer shell and mineral-wool acoustic fan enclosure
패널 두께30-60 mm double-wall cabinet and plenum sections
유효 폭900 - 1800 mm 명확한 작업 폭
최대 길이Up to 2400 mm overall cabinet height
강철 면면 두께1.0-1.5 mm SUS304 inner liner; 0.9-1.2 mm outer steel
코어 밀도80 - 120 kg / m 3 미네랄 울 음향 충전
열전도도 (Thermal Conductivity)0. 035 - 0. 045 W / (m · K) 음향 채우기
표면 마무리Fine-brushed SUS304 work zone with white epoxy-polyester outer coating
접합 유형Welded inner liner with negative-pressure plenum joints and gasketed supply and exhaust HEPA frames
사용자 정의의도된 응용 프로그램 및 운영 환경에 따라 사용할 수 있습니다.

사용 가능한 옵션

900-1800 mm work widthsRoom exhaust or canopy connection모터식 가죽Single-piece or segmented work tray