Prueba de integridad guante aislante
- In-Situ Pressure-Decay Test
- Inflatable Port Seal
- Wireless Portable Test Head
- Multiple Port Adapter Options
Introducción del producto
The glove integrity tester attaches to the isolator or RABS glove ring from inside the chamber. An inflatable circumferential seal fixes the lightweight test head in the port, then the instrument fills the glove and sleeve to the selected test pressure and records the decay over time. Battery power and wireless communication reduce hoses and cables across the barrier. Adapter sizes, test pressure and acceptance settings are matched to the actual glove material and port geometry.
Características clave
Pressure-decay measurement checks the glove and sleeve while installed in the port.
Battery power removes an external pneumatic hose from the chamber.
Wireless control can manage one or several test heads from a tablet or workstation.
Interchangeable adapters support different glove-ring diameters.
Preocupaciones del producto del cliente
Gloves are flexible barrier components that must be checked in their installed condition without creating a new leak around the test connection.
Aplicaciones típicas
Parámetros técnicos
Typical configurable construction ranges for an isolator glove integrity tester are listed below.
| Materiales Core | Anodized aluminum or SUS316L test head with engineering-polymer housing and inflatable silicone sealing ring |
|---|---|
| Panel de espesor | 3-8 mm instrument housing and adapter wall sections |
| Ancho efectivo | 200-400 mm compatible glove-port diameter |
| Longitud máxima | Up to 450 mm overall portable test-head length |
| Acero frente a espesor | 1.5-3.0 mm SUS316L adapter and protective components |
| Densidad Core | 1150-1300 kg/m³ for engineering-polymer housing components |
| Conductividad térmica | 0.20-0.35 W/(m·K) for the engineering-polymer housing |
| Superficie acabada | Smooth cleanroom-compatible polymer with anodized or electropolished metal adapters |
| Tipo conjunto | Tool-free inflatable circumferential port seal with interchangeable adapter ring and integrated pressure connection |
| Customization | Disponible según la aplicación prevista y el entorno operativo. |
Opciones disponibles
Planning a Isolator Glove Integrity Tester specification?
Envíenos sus dimensiones, condiciones de instalación, objetivos de rendimiento y acabado superficial requerido.

Inglés
Alemán
El árabe
Francés
Idioma indonesio
El japonés
Coreano
El malayo
El portugués
El ruso
EspañolCurrent
Tailandés
El turco
El vietnamita