클린룸 장비 및 무균 시설 필수품의 원스톱 공급업체국제 프로젝트 지원
Open mobile navigationClose mobile navigation
/ 제품 / 클린룸 유틸리티 시스템 / WFI Distribution Loop
클린룸 유틸리티 시스템

WFI 배포 루프

An insulated hygienic circulation skid for maintaining WFI flow and temperature through a hot storage-and-distribution network.
  • Insulated continuous hot circulation
  • Drainable low-hold-up user branches
  • Integrated thermal control and sanitization interface
견적 요청 제품 찾아보기

상품 소개

This WFI distribution module combines a hygienic circulation pump, insulated supply and return headers, thermal-control exchanger and short point-of-use branches on one open skid. Continuous return flow and measured temperature support the selected hot-storage or periodic thermal-sanitization strategy while keeping serviceable components accessible.

주요 기능

316L stainless product-contact tubing

Insulated supply and return headers

Hygienic centrifugal circulation pump

Drainable point-of-use branch geometry

Loop heat exchanger with isolated utility circuit

Temperature, pressure and conductivity monitoring

Automated steam or hot-water control interface

Local cabinet for pump and sanitization sequences

고객 제품 관심사

A WFI network must sustain circulation and controlled temperature without creating cool pockets, long branches or difficult-to-drain utility interfaces.

Return temperature falls at remote usersSize insulation, circulation flow and loop heat exchange from the full hydraulic and heat-loss profile.
Point-of-use branches cool during standbyUse compact continuously swept or recirculating branch arrangements with drainable hygienic valves close to the main header.
Thermal sanitization is unevenSequence pump speed and heat input from measured supply and return temperatures while verifying circulation through each defined path.
Utility leakage could contact WFISelect a leak-detectable double-tube-sheet exchanger or another isolated thermal interface suited to the operating strategy.

일반적인 응용 프로그램

Aseptic filling facilitiesParenteral formulation suitesVaccine and biologics manufacturingSterile equipment rinse utilitiesHigh-purity compounding areas

기술 매개변수

The loop skid is configured around storage temperature, return-temperature target, distribution length, peak demand and sanitization method.

핵심 재료316L stainless steel product-contact tubing, fittings, valve bodies, pump and exchanger surfaces
패널 두께Control cabinet and skid sheet components typically 1.5–3.0 mm
유효 폭Packaged hot-loop skid width typically 1200–2600 mm
최대 길이Packaged distribution module length typically up to 4500 mm
강철 면면 두께Stainless insulation cladding and cabinet facing typically 0.8–2.0 mm
코어 밀도316L stainless steel density approximately 8000 kg/m³
열전도도 (Thermal Conductivity)Typical hygienic insulation system approximately 0.035–0.045 W/(m·K)
표면 마무리Polished or electropolished wetted surfaces with cleanable satin stainless insulation cladding
접합 유형Orbital-welded main tubing with drainable sanitary service connections
사용자 정의의도된 응용 프로그램 및 운영 환경에 따라 사용할 수 있습니다.

사용 가능한 옵션

Duty and standby pumpsDouble-tube-sheet loop heat exchangerSteam or hot-water heating interfaceContinuously swept point-of-use subloopsReturn conductivity and TOC monitoringAutomated thermal-sanitization sequenceInsulated valve jacketsRemote data and alarm integration

Planning a WFI Distribution Loop specification?

치수, 설치 조건, 성능 목표 및 필요한 표면 마감을 보내십시오.

우리 팀과 대화하십시오.
저희에게 이메일 whatsapp WeChat

제품

뉴스

회사 소개

문의하기