A molecular filter using acid-treated activated carbon or engineered pellets to capture ammonia, amines and other basic gaseous contaminants.
  • Base-Gas Selectivity
  • Acid-Impregnated Media
  • High Adsorbent Mass
  • Contained Pellet Bed
Solicite un presupuesto. Browse productos

Introducción del producto

The alkaline gas molecular filter is built around an acid-impregnated porous adsorbent that neutralizes ammonia and related basic vapours. Granules are densely packed between rigid mesh faces so the air stream crosses a consistent reaction bed. The cell can be installed in a standard side-access housing or a safe-change system downstream of particulate prefiltration. Separate basic-gas media prevents the premature exhaustion that occurs when general-purpose carbon is exposed to high ammonia loading.

Características clave

Acidic impregnation promotes chemisorption of ammonia and light amines.

Deep packed media provides longer residence time than a thin carbon sheet.

Sealed perimeter gasketing directs all air through the active bed.

Low-dust screens protect downstream HEPA stages from pellet abrasion.

Preocupaciones del producto del cliente

Ammonia and amines are difficult to retain on untreated carbon and can quickly spread through recirculating ventilation.

General odour carbon shows early breakthrough when ammonia concentration rises.Acid-impregnated media reacts with alkaline molecules and provides higher usable capacity.
A thin pleated filter lacks enough media mass for continuous base-gas exposure.The deep granular bed increases adsorbent quantity and contaminant contact time.
Pellet movement creates voids after shipment.Retaining grids and bed compression keep the media evenly distributed across the face.
Combining acid and base control in one bed wastes selective capacity.A dedicated alkaline-gas stage can be arranged separately from acid- and VOC-removal stages.

Aplicaciones típicas

Animal research facilitiesFertilizer and chemical laboratoriesBattery process ventilationWaste treatment roomsAmine-handling production areas

Parámetros técnicos

Representative dimensions and materials for a base-gas molecular filter are shown below.

Materiales CoreAcid-impregnated activated carbon or engineered alumina pellets for ammonia and amines
Panel de espesor150–292 mm overall filter depth
Ancho efectivo305–610 mm nominal cell width
Longitud máximaUp to 610 mm nominal cell height
Acero frente a espesor0.8–1.2 mm galvanized or 304 stainless-steel enclosure
Densidad Core480–700 kg/m³ packed impregnated-carbon bulk density
Conductividad térmica0.10–0.22 W/(m·K) for the granular adsorbent bed
Superficie acabadaMill-galvanized, brushed stainless steel or chemical-resistant powder coat
Tipo conjuntoSealed box cell with dual perforated retainers and moulded neoprene face gasket
CustomizationDisponible según la aplicación prevista y el entorno operativo.

Opciones disponibles

Ammonia-specific media gradeDisposable or refillable cellFull-size or half-size formatBIBO-compatible gasket arrangement