Class III 생물 안전 캐비닛
- Gas-Tight Negative-Pressure Enclosure
- Four Fixed Gauntlet Ports
- Single-HEPA Supply and Double-HEPA Exhaust
- Interlocked Material Pass-Through
상품 소개
The Class III Biosafety Cabinet is a completely enclosed, ventilated glove-operated workstation for laboratory processes that require a physical barrier around the manipulation zone. Unlike an open-front Class II cabinet, the selected four-port configuration has one fixed sealed observation window and four mechanically attached neoprene gauntlets, so routine hand access does not create an open working aperture. Room air enters through one dedicated HEPA stage, while cabinet air leaves through two HEPA stages in series and a hard-duct exhaust collar; the connected exhaust system is selected to keep the work enclosure below the surrounding room pressure. A double-ended side pass-through uses a one-door-at-a-time interlock to move closed materials without opening the main chamber. The satin AISI 304 work zone is contained within a white powder-coated steel shell and supported on a mobile leveling stand. A local display, differential-pressure gauge and separate green-red status indicators make operating and alarm conditions visible. Final pressure, exhaust capacity, filter arrangement, transfer method and facility connection are configured for the site's risk assessment rather than inferred from the cabinet name alone.
주요 기능
One filtered supply path and two serial exhaust filter stages are paired with a dedicated duct collar so the facility exhaust can maintain inward leakage and negative cabinet pressure.
The integral double-ended transfer chamber uses opposing gasketed doors and a one-door-at-a-time interlock; it is a containment transfer route, not an autoclave or automatic sterilizer.
A local pressure gauge, control display and separate normal-alarm indicators provide direct feedback without opening the cabinet or relying only on a remote building interface.
고객 제품 관심사
A Class III cabinet must keep the work zone physically closed, maintain a reliable filtered pressure path and provide a material route that does not bypass the containment boundary.
일반적인 응용 프로그램
기술 매개변수
Representative configurable construction values for the illustrated four-gauntlet, negative-pressure Class III cabinet are listed below.
| 핵심 재료 | Fully welded gas-tight AISI 304 stainless-steel work enclosure with a white powder-coated electro-galvanized external shell, fixed safety glazing, neoprene gauntlets and gasketed HEPA housings |
|---|---|
| 패널 두께 | 1.2 mm nominal AISI 304 work-zone panel |
| 유효 폭 | 1430 mm nominal internal work-zone width for the selected four-port configuration |
| 최대 길이 | Up to 1995 mm nominal overall equipment length including the integral side pass-through in the selected configuration |
| 강철 면면 두께 | 1.5 mm nominal powder-coated steel external facing |
| 코어 밀도 | 7900 kg/m³ for representative AISI 304 stainless-steel components |
| 열전도도 (Thermal Conductivity) | 15 W/(m·K) for representative AISI 304 stainless steel at room temperature |
| 표면 마무리 | White powder-coated exterior, neutral silver satin AISI 304 work zone, colorless clear glazing, charcoal-black gauntlets and seals, dark-blue display, green and red status lenses |
| 접합 유형 | Fully welded gas-tight enclosure with a fixed sealed viewing panel, four mechanically clamped gauntlet ports, an interlocked double-ended pass-through and a hard-connected single-supply/double-exhaust HEPA path |
| 사용자 정의 | 의도된 응용 프로그램 및 운영 환경에 따라 사용할 수 있습니다. |