ST-LPW16-005
The 5 L/h model is the lowest-output choice for small, occasional ultrapure-water preparations; the standard pressure tank is 5 gal.
Cleanroom Pure Water Systems
| Brand | ST |
|---|---|
| Public model series | ST-LPW16 |
| Feed-water requirements | municipal tap water; TDS ≤200 ppm, pressure 1.0–5.0 kg/cm², temperature 5–45°C |
| Main unit dimensions | H × W × D: 512 × 390 × 445 (mm) |
| Unit weight | 30-50 kg |
| Power supply and rated power | AC 220 V/50 Hz (power 30-50 W) |
| Storage tank | supplied with 5 gal pressure tank (other sizes optional) |
| Dispensing rate | 1.8-2.0 L/min |
| RO water quality | conductivity ≤1-5 µS/cm (online monitoring) |
| Ultrapure water quality | resistivity 18.2 MΩ·cm at 25°C (online monitoring, GB/T 6682-2008) |
| Microbial level | ≤1 CFU/mL |
| Particle count | ≤ 1 particle/mL |
| Heavy-metal ions | ≤0.1 ppb |
Customization availableSpecifications can be customized to your requirements.
Specify the quantity and configuration in your inquiry.
Laboratory Ultrapure Water System – ST-LPW16 Series supplies monitored RO and ultrapure water in a compact housing for laboratories needing both routine preparation and a higher-purity final water source. The stated heavy-metal value adds a useful selection criterion for ion-sensitive work.
The standard 5 gal tank buffers intermittent collection while the output choices set replenishment capacity. The cabinet stays compact; the larger output options are useful when batches are repeated more frequently, not because each litre is a different water grade.
MODELS AND DIMENSIONS
| Model | Production rate |
|---|---|
| ST-LPW16-005 | 5 L/h at 25°C |
| ST-LPW16-010 | 10 L/h at 25°C |
| ST-LPW16-015 | 15 L/h at 25°C |
| ST-LPW16-020 | 20 L/h at 25°C |
| ST-LPW16-040 | 40 L/h at 25°C |
| ST-LPW16-060 | 60 L/h at 25°C |
| ST-LPW16-100 | 100 L/h at 25°C |
Production-rate ratings are stated at the temperature shown in the table.
MODEL DIFFERENCES
The 5 L/h model is the lowest-output choice for small, occasional ultrapure-water preparations; the standard pressure tank is 5 gal.
A 10 L/h rating provides twice the recovery capacity for a bench where small preparations recur throughout the day.
The 15 L/h version is an intermediate choice when 10 L/h leaves too little replenishment time but 20 L/h is unnecessary.
The 20 L/h model supports a more regular batch schedule without changing the standard pressure-tank size.
At 40 L/h, the production step is useful when several batches are prepared together rather than separated by long idle periods.
The 60 L/h option gives more replenishment during sustained preparation, with the same listed water-quality specification.
The 100 L/h version is the highest-output configuration; its standard 5 gal tank does not grow with production, so burst volume and recovery demand remain separate choices.
FEATURES AND BENEFITS
The heavy-metal specification supplements bulk resistivity when assessing preparation for sensitive measurements.
Small production options suit spaced draws; larger options reduce the time needed to restore the water reserve.
APPLICATIONS
Place a monitored water point near repeated preparation and rinsing tasks, with each outlet selected for the task’s required water quality.
SYSTEM SELECTION
Include the supplied heavy-metal value in the preparation-water decision when trace ionic contamination matters to the method.
Plan the external tank beside the compact unit, then select the output step around the largest collection and the time before water is needed again.
PROCUREMENT / INQUIRY
Tell us the quantity and configuration you need. This product will be included automatically.
Fields marked * are required.