ST / INTEGRATED CLEANROOM PROCUREMENTEquipment / Consumables / Supporting facilities
PRODUCT / TECHNICAL DATA Clean Laboratory Equipment & SuppliesSpecifications / Configurations / Procurement

Clean Laboratory Equipment & Supplies

All-Steel Horizontal Laminar Flow Clean Bench

Product specifications

BrandST
ConstructionPowder-coated cold-rolled steel
AirflowHorizontal: rear to front
Air intake and filtrationTop air intake; pre-filter and HEPA filter
HEPA filter mediaMini-pleat, separator-free ultrafine glass fibre
Cleanliness classClass 100 for particles ≥0.5 μm (US Federal Standard 209E)
Filtration efficiency99.995% at 0.5 μm
Average air velocity0.25–0.45 m/s
Fan speed controlTwo speeds: high / low
Settling colony count≤0.5 CFU per dish per hour; Ø90 mm settle plate
Vibration half-amplitude≤5 μm in X, Y and Z directions
Illuminance≥300 lx
UV wavelength254 nm
Maximum power consumption300 W / 600 W, according to configuration
Noise≤62 dB
Power supplyAC 220 V, single-phase, 50 Hz
Additional open configurationsST-CS111-05 single-person; ST-CS111-06 two-person
CustomisationDimensions, working access and configuration

Customization availableSpecifications can be customized to your requirements.

Request a quotation

Specify the quantity and configuration in your inquiry.

All-Steel Horizontal Laminar Flow Clean Bench passes filtered air from the rear of the enclosure across the work towards the open front. It provides a protected working area for clean handling, assembly and inspection where airborne dust can lead to repeated cleaning or rejected surfaces.

The rear filter makes the position of each object important: keep the cleanest item upstream and avoid putting hands or bulky tools between the filter and that item.

MODELS AND DIMENSIONS

Models and working dimensions

Dimensions are length × depth × height in mm. Lighting and UV columns show lamp power and quantity.
ModelFormatOperators / accessWork area (mm)Overall (mm)LightingUV lamp
ST-CS111-01Benchtop1 / single-sided490 × 440 × 480620 × 520 × 83010 W × 110 W × 1
ST-CS111-02Benchtop1 / single-sided680 × 440 × 480820 × 520 × 83015 W × 115 W × 1
ST-CS111-03Floor-standing1 / single-sided700 × 440 × 560850 × 520 × 160020 W × 120 W × 1
ST-CS111-04Floor-standing2 / single-sided1150 × 440 × 5601290 × 520 × 160030 W × 130 W × 1

MODEL DIFFERENCES

Which configuration fits the task

490 mm benchtop workspace

The compact horizontal-flow area suits a small part positioned close to the rear clean-air source.

680 mm benchtop workspace

Extra working length accommodates a wider part or a small set of tools on an existing support.

700 mm floor-standing workspace

This dedicated one-operator station places the horizontal-flow enclosure at its own working position.

1150 mm floor-standing workspace

The longer two-operator format gives more lateral room; arrange each task so objects do not block one another’s upstream air.

FEATURES AND BENEFITS

How it helps

Direct rear-to-front supply

Place critical materials closer to the clean supply and keep other objects from interrupting the airflow before it reaches them.

Steel cabinet construction

The coated steel body provides a rigid enclosure for repeated preparation, inspection and assembly work.

Open working access

Single- and two-person open configurations accommodate work that needs direct access across the front.

APPLICATIONS

Where it is used

01

Electronics assembly and inspection

Sweep filtered air across clean components while positioning tools outside the upstream air path.

02

Optical component handling

Limit airborne dust deposition during non-hazardous lens and optical-part inspection.

03

Biology and plant tissue-culture laboratories

Support clean handling of non-hazardous samples and plant cultures with the workspace arranged for rear-to-front flow.

04

Pharmaceutical research laboratories

Provide a clean local work area for non-hazardous materials and research sample preparation.

SELECTION GUIDANCE

Choosing the configuration

Match width to the task

Use an HD benchtop unit for smaller parts, or a single- or two-person workstation for wider work.

Keep the rear supply clear

Avoid placing tall containers or equipment between the filter outlet and exposed materials.

PRACTICAL QUESTIONS

Questions about this product

What does the clean bench protect?

It protects the work from airborne contamination. Use it for non-hazardous materials; it does not provide personnel protection or replace a biosafety cabinet or chemical fume hood.

How should the UV lamp be used?

Use UV treatment only when the bench is unoccupied, with eyes and skin protected from exposure. Switch the UV lamp off before working; UV treatment does not replace surface cleaning.

PROCUREMENT / INQUIRY

Request a
product quotation.

Tell us the quantity and configuration you need. This product will be included automatically.

Fields marked * are required.

YOUR PRODUCTAll-Steel Horizontal Laminar Flow Clean BenchReview specifications
Useful details for this quotation

Show the working layout so the horizontal airflow path stays clear around the exposed product.

  • Work and material dimensions
  • Worktop width
  • Site power supply
  • Clear space in front of the bench

We use your contact details and requirements to review your inquiry and prepare a quotation. Sending this form does not place an order.

All-Steel Horizontal Laminar Flow Clean Bench — product view

YOUR PURCHASING BRIEF

Inquiry list

Collect products here, then send your requirements in one inquiry.

Continue to inquiry ↗

This list is saved in this browser. Adding a product does not place an order.